Z-0121-2008

The Brilliance CT Big Bore is a whole body computed tomography X-Ray system. The malfunctioning software...

Philips North America LLC / initiated 2007-09-27 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of The Brilliance CT Big Bore is a whole body computed tomography X-Ray system. The malfunctioning software application is the Tumor Localization (Tumor LOC)... on . The reason FDA records is "The maximum intensity projection (MIP) images produced from respiratory gated images by the Tumor Localization (Tumor LOC) software application on Brilliance Big Bore CT scanners will be labeled incorrectly for certain...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0121-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
The Brilliance CT Big Bore is a whole body computed tomography X-Ray system. The malfunctioning software application is the Tumor Localization (Tumor LOC) application.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The maximum intensity projection (MIP) images produced from respiratory gated images by the Tumor Localization (Tumor LOC) software application on Brilliance Big Bore CT scanners will be labeled incorrectly for certain patient orientations and view conventions when saved for later viewing.
Root cause tag
software
FDA root cause
Software design
Action
Notification letters were sent to the installed customer base informing them of the hazard on 10/1/2007. A mandatory corrective software release due out mid October 2007, will be available at no cost to the customer. The software will be installed by Philips Field Service Engineers.
Quantity in commerce
206 systems
Distribution
Worldwide, including USA, Canada, France, Finland, Korea, Australia, Italy, Switzerland, China, Austria, Thailand, Malaysia, Denmark, Norway, Jordan, Netherlands, England, South Africa, Spain, Germany, Bahrain, Belgium, New Zealand, Puerto Rico, South Korea, Saudi Arabia, Czech Republic, Sweden, Israel, and Lithuania
Product codes and lots
All Brilliance CT Big Bore Scanners with Tumor Localization software versions v2.2.1, v2.2.2, & v2.2.5. This software version is included on Brilliance CT Big Bore Scanners, Model Number 728243, and Serial Numbers 7001 - 7215.