Z-0245-2008

Clearview hCG Combo Waived for urine & Moderate for serum, Wampole Laboratories, ABI List No/Product No...

Applied Biotech Inc / initiated 2006-09-28 / Terminated / root cause: manufacturing process

Applied Biotech Inc began a recall of Clearview hCG Combo Waived for urine & Moderate for serum, Wampole Laboratories, ABI List No/Product No. 6026KCV; Contents 30 Individually pouched test... on . The reason FDA records is "Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0245-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Applied Biotech Inc
Product
Clearview hCG Combo Waived for urine & Moderate for serum, Wampole Laboratories, ABI List No/Product No. 6026KCV; Contents 30 Individually pouched test cassettes with disposable pipettes, 1 Directional Insert; Inverness Medical Professional Diagnostics Group; Marketed by Wampole Laboratories, 2 Research Way, Princeton NJ 08540 USA
Product code
JHI Visual, Pregnancy Hcg, Prescription Use
Reason for recall
Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the product.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
A Product Recall Notification dated 9/28/2006, was provided via fax, email or direct mail to all primary consignees. Customers were instructed to block all stock of affected lots in their warehouses, contact their primary consignees and advise them to discontinue use immediately and dispose of any unused materials; using a fax back form to communicate to the recalling firm the quantity blocked and disposed of so that arrangements could be made for providing replacement product
Quantity in commerce
3,238 Kits (97,140 tests)
Distribution
Worldwide: USA and Canada
Product codes and lots
Model Number 6026KCV; ABI Lot Code: 87494