Z-0330-2008
Allura Xper FD20/10, Release 1.2, Philips Medical Systems, Bothell, WA 98021
Philips North America LLC / initiated 2007-09-24 / Terminated / root cause: design
Philips Medical Systems North America Co. Phillips began a recall of Allura Xper FD20/10, Release 1.2, Philips Medical Systems, Bothell, WA 98021 on . The reason FDA records is "Inaccurate anatomical measurements when using the optional SmartMask feature". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0330-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Allura Xper FD20/10, Release 1.2, Philips Medical Systems, Bothell, WA 98021
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Inaccurate anatomical measurements when using the optional SmartMask feature
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The recalling firm sent customers an Urgent Device Correction notification letter dated 09/21/07 via certified mail - return receipt requested. This notification provided a short term work around for the device defect. For a permanent fix, the recalling firm is developing software update to check for mismatch at the start of a subtracted fluoroscopy run, and will contact customers to schedule an install. Installation is expected to begin 10/2007 and be completed 06/2008.
- Quantity in commerce
- 53 units
- Distribution
- Nationwide
- Product codes and lots
- Serial numbers affected: 18, 19, 22, 23, 24, 32, 33, 34, 36, 38, 52, 54, 56, 67, 68, 69, 71, 73, 74, 82, 84, 94, 99, 100, 102, 103, 106, 108, 112, 113, 116, 124, 125, 126, 131, 133, 141, 142, 148, 154, 157, 173, 174, 176, 178, 187, 191, 192, 197, 199, 212, and 691.