Z-0336-2008
Axya Anchor, 5.0 mm Titanium in Super Kit, Catalog Number: CAT1270, Axya Medical, Inc., Beverly, MA 01915
TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging
Axya Medical began a recall of Axya Anchor, 5.0 mm Titanium in Super Kit, Catalog Number: CAT1270, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0336-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- Axya Anchor, 5.0 mm Titanium in Super Kit, Catalog Number: CAT1270, Axya Medical, Inc., Beverly, MA 01915
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
- Quantity in commerce
- 208 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 203415, 203445, 203588, 203604, 203739, 203742, 203791, 203798, 203820, 203823, 203834, 203863, 203905, 203915, 203994, 204025, 204126, 204176, 204187, 204199, 204209, 204239, 204246, 204277, 204302, 204308, 204350, 204355, 204390, 204459,204482, 204511, 204513, 204516, 204521, 204546, 204626, 204642, 204646, 204651,204673, 204692, 204721, 204736, and 204781