Z-0339-2008
Axya Bone Anchor 5.0 mm Bioabsorbable with Nylon Suture Catalog Number: CAT1234, Axya Medical, Inc., Beverly...
TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging
Axya Medical began a recall of Axya Bone Anchor 5.0 mm Bioabsorbable with Nylon Suture Catalog Number: CAT1234, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0339-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- Axya Bone Anchor 5.0 mm Bioabsorbable with Nylon Suture Catalog Number: CAT1234, Axya Medical, Inc., Beverly, MA 01915
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
- Quantity in commerce
- 904 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 203816, 203856, 203862, 203900, 203912, 204572, 204574, 204596, 204613, 204619, and 204715
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
