Z-0344-2008

Axya Bone Anchor, 5.0 mm Bioabsorbable in Super Kit, Catalog Number: CAT1271, Axya Medical, Inc., Beverly, MA...

TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging

Axya Medical began a recall of Axya Bone Anchor, 5.0 mm Bioabsorbable in Super Kit, Catalog Number: CAT1271, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0344-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
TORNIER Inc.
Product
Axya Bone Anchor, 5.0 mm Bioabsorbable in Super Kit, Catalog Number: CAT1271, Axya Medical, Inc., Beverly, MA 01915
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
Quantity in commerce
329 units
Distribution
Nationwide
Product codes and lots
Lot Numbers: 203864, 203957, 203973, 204024, 204035, 204111, 204127, 204148, 204177, 204183, 204228, 204238, 204243, 204247, 204248, 204262, 204278, 204291, 204303, 204306, 204307, 204319, 204321, 204342, 204351, 204387, 204417, 204422, 204445, 204460,204461, 204474, 204475, 204478, 204488, 204504, 204512, 204515, 204522, 204566,204568, 204569, 204675, 204688, 204691, 204722, 204723, 204782, 204789, 204796,204842, 204843, 204866, 204894, 204955, 204969, 205000, 205053, 205110 and 205658