Z-0345-2008

Axya Bone Anchor 3.0 mm Bioabsorbable in Super Kit Catalog Number: CAT1273, Axya Medical, Inc., Beverly, MA...

TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging

Axya Medical began a recall of Axya Bone Anchor 3.0 mm Bioabsorbable in Super Kit Catalog Number: CAT1273, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0345-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
TORNIER Inc.
Product
Axya Bone Anchor 3.0 mm Bioabsorbable in Super Kit Catalog Number: CAT1273, Axya Medical, Inc., Beverly, MA 01915
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
Quantity in commerce
59 units
Distribution
Nationwide
Product codes and lots
Lot Numbers: 204477, 204483, 204489, 204514, 204520, 204524, 204547, 204554, 204565, 204573, 204653, 204662, 204674, 204724, 204956, 204970 and 205001