Z-0345-2008
Axya Bone Anchor 3.0 mm Bioabsorbable in Super Kit Catalog Number: CAT1273, Axya Medical, Inc., Beverly, MA...
TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging
Axya Medical began a recall of Axya Bone Anchor 3.0 mm Bioabsorbable in Super Kit Catalog Number: CAT1273, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0345-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- Axya Bone Anchor 3.0 mm Bioabsorbable in Super Kit Catalog Number: CAT1273, Axya Medical, Inc., Beverly, MA 01915
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
- Quantity in commerce
- 59 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 204477, 204483, 204489, 204514, 204520, 204524, 204547, 204554, 204565, 204573, 204653, 204662, 204674, 204724, 204956, 204970 and 205001