Z-0347-2008

Arthocare 3mm ParaFix Anchor (3mm x 9mm) -single strand of #2 braided suture,Part Number: 22-5002, Axya...

TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging

Axya Medical began a recall of Arthocare 3mm ParaFix Anchor (3mm x 9mm) -single strand of #2 braided suture,Part Number: 22-5002, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0347-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
TORNIER Inc.
Product
Arthocare 3mm ParaFix Anchor (3mm x 9mm) -single strand of #2 braided suture,Part Number: 22-5002, Axya Medical, Inc., Beverly, MA 01915
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
Quantity in commerce
321 units
Distribution
Nationwide
Product codes and lots
Lot Numbers: 203144, 203228, 203333 and 203719