Z-0348-2008

Arthrocare 3mm ParaFix Anchor, single strand of Magnum Wire,Part Number: 22-5010, Axya Medical, Inc...

TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging

Axya Medical began a recall of Arthrocare 3mm ParaFix Anchor, single strand of Magnum Wire,Part Number: 22-5010, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0348-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
TORNIER Inc.
Product
Arthrocare 3mm ParaFix Anchor, single strand of Magnum Wire,Part Number: 22-5010, Axya Medical, Inc., Beverly, MA 01915
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
Quantity in commerce
338 units
Distribution
Nationwide
Product codes and lots
Lot Numbers: 203734, 204097, 204098, 204223 and 204908