Z-0351-2008
Arthrocare 3.0mm ParaSorb, Preloaded with 2 MagnumWire sutures (1 white/1 co-braid), Part Number: 22-5030...
TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging
Axya Medical began a recall of Arthrocare 3.0mm ParaSorb, Preloaded with 2 MagnumWire sutures (1 white/1 co-braid), Part Number: 22-5030, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0351-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- Arthrocare 3.0mm ParaSorb, Preloaded with 2 MagnumWire sutures (1 white/1 co-braid), Part Number: 22-5030, Axya Medical, Inc., Beverly, MA 01915
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
- Quantity in commerce
- 1,708 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 204007 204008 204016 204017 204045 204057 204058 204069 204072 204530 204531 204532 204548 204563 204564 204570 204571 and 204588