Z-0353-2008
Arthrocare 5.5mm ParaSorb, pre-loaded with 2 MagnumWire sutures (1 white/1 co-braid) with needle Part Number...
TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging
Axya Medical began a recall of Arthrocare 5.5mm ParaSorb, pre-loaded with 2 MagnumWire sutures (1 white/1 co-braid) with needle Part Number: 22-5055N, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0353-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- Arthrocare 5.5mm ParaSorb, pre-loaded with 2 MagnumWire sutures (1 white/1 co-braid) with needle Part Number: 22-5055N, Axya Medical, Inc., Beverly, MA 01915
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
- Quantity in commerce
- 337 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 205278 205280 205281 205310 205311 205312 205319 205329 205407 205450 205451 205503 205504 205505 205506 205569 205570 205571 205593 205601 and 205631