Z-0354-2008
Arthrocare 6.5 mm ParaSorb, Preloaded with 2 MagnumWire sutures (1 white/1 co-braid), Part Number: 22-5065...
TORNIER Inc. / initiated 2007-09-27 / Terminated / root cause: packaging
Axya Medical began a recall of Arthrocare 6.5 mm ParaSorb, Preloaded with 2 MagnumWire sutures (1 white/1 co-braid), Part Number: 22-5065, Axya Medical, Inc., Beverly, MA 01915 on . The reason FDA records is "Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0354-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- Arthrocare 6.5 mm ParaSorb, Preloaded with 2 MagnumWire sutures (1 white/1 co-braid), Part Number: 22-5065, Axya Medical, Inc., Beverly, MA 01915
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
- Quantity in commerce
- 1,538 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 203767 203769 203809 203829 203839 203940 203941 203942 203953 203965 203966 203967 203968 204010 204018 204019 204130 204236 204242 204250 204251 204252 204284 204285 204286 204292 204293 204294 204295 204314 204315 204316 204356 204357 204358 204359 204360 204361 204368 204377 204379 204380 204400 204401 204412 204433 204434 204437 204438 204609 204610 204611 204624 204657 204667 204668 and 204669