Z-0374-2008

Abbott Point of Care, i-STAT EC8+ Cartridge; (NA K CL Glu pH PCO 2 UN/Urea Hct); List No: 06F04-01; Part No...

Abbott Medical / initiated 2007-10-18 / Terminated / root cause: packaging

Abbott Point of Care Inc. began a recall of Abbott Point of Care, i-STAT EC8+ Cartridge; (NA K CL Glu pH PCO 2 UN/Urea Hct); List No: 06F04-01; Part No: 12500; Abbott Point of Care Inc., East Windsor, NJ... on . The reason FDA records is "Difficult to Close -- Certain lots of i-STAT cartridges are difficult to close, will not close or do not stay closed". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0374-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott Point of Care, i-STAT EC8+ Cartridge; (NA K CL Glu pH PCO 2 UN/Urea Hct); List No: 06F04-01; Part No: 12500; Abbott Point of Care Inc., East Windsor, NJ 08520, USA; Product of Canada.
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
Difficult to Close -- Certain lots of i-STAT cartridges are difficult to close, will not close or do not stay closed.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Abbott Point of Care sent and Urgent Notice to all i-STAT Customers (US and Canadian consignees) notifying them that certain lots of i-STAT cartridges have snap closures that are difficult to close, will not close or do not stay closed. The customers were asked to review the lot numbers on the enclosure and, if the facility has any cartridges from the lot numbers listed, they should remove from service immediately.
Distribution
Worldwide Distribution- including USA and country of Canada.
Product codes and lots
Catalog Number: 06F04-01; 06F04-02 Lot Numbers: A07110 and A07153.