Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0269-2008

Siemens ECAT ACCEL, Catalog No. 3554479, PET/CT Scanner, Manufactured by Siemens Medical Solutions USA, Inc...

Siemens Medical Solutions USA, Inc. / initiated 2007-07-23 / Terminated / root cause: software

Siemens Medical Solutions USA Inc began a recall of Siemens ECAT ACCEL, Catalog No. 3554479, PET/CT Scanner, Manufactured by Siemens Medical Solutions USA, Inc., 810 Innovation Drive, Knoxville, TN 37932 USA... on . The reason FDA records is "Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0269-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens ECAT ACCEL, Catalog No. 3554479, PET/CT Scanner, Manufactured by Siemens Medical Solutions USA, Inc., 810 Innovation Drive, Knoxville, TN 37932 USA, TEL: 865-218-2534
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
Root cause tag
software
FDA root cause
Software design
Action
Siemans sent an Urgent Device Correction letter, dated June 15, 2007, to all affected consignees informing the user of the problem and requesting that they discontinue the practice of running all transmissions followed by all emissions until a corrective software patch is installed. The corrective software patch is expected to be installed by October 2007.
Quantity in commerce
105 units
Distribution
Worldwide Distribution - USA and Countries of Argentina, Austria, Belgium, Brazil, Canada, Switzerland, Chile, China, Czech Republic, Germany, Denmark, Spain, Finland, France, Great Britain, Italy, Japan, South Korea, Netherlands, Norway, Saudi Arabia, Sweden, Slovak Republic, Turkey, and Taiwan.
Product codes and lots
Serial Numbers: 101014, 1303, 1305, 1309, 1310, 1311, 1314, 301089, 301092, 301111, 3600211-00-001003, 3600211-00-001041, 3600211-00-001110, 3600211-00-001122, 3600211-00-0101001, 3600211-00-0101004, 3600211-00-0101005, 3600211-00-0101006, 3600211-00-0101008, 3600211-00-0101012, 3600211-00-0101015, 3600211-00-0101016, 3600211-00-0101017, 3600211-00-0201007, 3600211-00-0201022, 3600211-00-0201023, 3600211-00-0201024, 3600211-00-0201025, 3600211-00-0201026, 3600211-00-0201027, 3600211-00-0201029, 3600211-00-0201031, 3600211-00-0201032, 3600211-00-0201033, 3600211-00-0201035, 3600211-00-0201036, 3600211-00-0201037, 3600211-00-0201038, 3600211-00-0201039, 3600211-00-0201040, 3600211-00-0201041, 3600211-00-0201042, 3600211-00-0201043, 3600211-00-0201045, 3600211-00-0201049, 3600211-00-0201050, 3600211-00-0201051, 3600211-00-0201052, 3600211-00-0201053, 3600211-00-0201054, 3600211-00-0201055, 3600211-00-0201056, 3600211-00-0201057, 3600211-00-0201058, 3600211-00-0201060, 3600211-00-0201061, 3600211-00-0201062, 3600211-00-0201063, 3600211-00-0201064, 3600211-00-0201065, 3600211-00-0201066, 3600211-00-0201067, 3600211-00-0201069, 3600211-00-0201070, 3600211-00-0201071, 3600211-00-0201072, 3600211-00-0201079, 3600211-00-0201080, 3600211-00-0201084, 3600211-00-0201090, 3600211-00-0201091, 3600211-00-0201093, 3600211-00-0201095, 3600211-00-0201103, 3600211-00-0201104, 3600211-00-0201106, 3600211-00-0201115, 3600211-00-0201120, 3600211-00-0301073, 3600211-00-0301075, 3600211-00-0301076, 3600211-00-0301077, 3600211-00-0301078, 3600211-00-0301083, 3600211-00-0301085, 3600211-00-0301086, 3600211-00-0301087, 3600211-00-0301094, 3600211-00-0301096, 3600211-00-0301098, 3600211-00-0301099, 3600211-00-0301100, 3600211-00-0301102, 3600211-00-0301107, 3600211-00-0301108, 3600211-00-0301109, 3600211-00-0401112, 3600211-00-0401113, 3600211-00-0401114, 3600211-00-0401116, 3600211-00-0401117, 3600211-00-201034, 3600211-00-201046, 3600211-00-301097, 3600211-00-301105, 401118