Siemens ECAT ART, Catalog No.. 3545584, PET/CT Scanner, Manufactured by Siemens Medical Solutions USA, Inc...
Siemens Medical Solutions USA Inc began a recall of Siemens ECAT ART, Catalog No.. 3545584, PET/CT Scanner, Manufactured by Siemens Medical Solutions USA, Inc., 810 Innovation Drive, Knoxville, TN 37932 USA... on . The reason FDA records is "Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0270-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens ECAT ART, Catalog No.. 3545584, PET/CT Scanner, Manufactured by Siemens Medical Solutions USA, Inc., 810 Innovation Drive, Knoxville, TN 37932 USA, TEL: 865-218-2534
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemans sent an Urgent Device Correction letter, dated June 15, 2007, to all affected consignees informing the user of the problem and requesting that they discontinue the practice of running all transmissions followed by all emissions until a corrective software patch is installed. The corrective software patch is expected to be installed by October 2007.
- Quantity in commerce
- 43 units
- Distribution
- Worldwide Distribution - USA and Countries of Argentina, Austria, Belgium, Brazil, Canada, Switzerland, Chile, China, Czech Republic, Germany, Denmark, Spain, Finland, France, Great Britain, Italy, Japan, South Korea, Netherlands, Norway, Saudi Arabia, Sweden, Slovak Republic, Turkey, and Taiwan.
- Product codes and lots
- Serial Numbers: 3600066-00-0001001, 3600066-00-0001002, 3600066-00-0001005, 3600066-00-0001007, 3600066-00-0001008, 3600066-00-0001010, 3600066-00-0001013, 3600066-00-0001014, 3600078-00-0001003, 3600082-00-0001002, 3600084-00_0001005, 3600084-00-0001005, 3600090-00-0001001, 3600090-00-0001002, 3600090-00-0001003, 3600090-00-0201005, 3600090-00-0201006, 3600099-00-001001, 3600099-00-001006, 3600099-00-001009, 3600099-00-0201002, 3600099-00-0201003, 3600099-00-0201004, 3600099-00-0201005, 3600099-00-0201007, 3600099-00-0201008, 3600099-00-0201010, 3600099-00-0201012, 3600099-00-0301013, 3600099-00-0301014, 3600099-00-0301016, 3600099-00-0301017, 3600099-00-0301018, 3600099-00-0301019, 3600099-00-0301020, 3600099-00-0301021, 3600099-00-0301022, 3600099-00-0301023, 3600099-00-0301024, 3600099-00-0301025, 3600099-00-0301026, 3600099-00-0301027, 3600099-00-0301028