Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC...
Nicolet Biomedical Div of Viasys Healthcare began a recall of Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between... on . The reason FDA records is "Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0126-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system), released and installed between December 2004 and March 2007 (Axiomtek Panel PC Model Number: 2147-671) VIASYS NeuroCare Part Numbers: 842-673800, 842-673900, 982A0421 and 982A0422. Manufactured, designed, developed and marketed by VIASYS NeuroCare Inc., Madison, WI 53711
- Product code
- GWQ Full-Montage Standard Electroencephalograph
- Reason for recall
- Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Certified direct mail delivery of URGENT MEDICAL DEVICE FIELD CORRECTION (RECALL) were made to consignees on 9/11/07 by Nicolet Biomedical Div of Viasys Healthcare. The letter describes the product and problem. VIASYS requests that customers cease use of the video option of the device until the monitor can be replaced by VIASYS authorized personnel. Letter also requests consignees to fax or mail back the response form.
- Quantity in commerce
- 33
- Distribution
- Worldwide Dsitribution - USA, Germany, Netherlands, Kuwait, Japan, Ireland, and China.
- Product codes and lots
- ON040060, ON050120, ON050130, ON050168, ON050186, ON050197, ON050199, ON050203, ON050206, ON050212, ON050220, ON050231, ON060232, ON060243, ON060254, ON060256, ON060257, ON060269, ON060270, ON060272, ON060275, ON060276, ON060277, ON060278, ON060279, ON060280, ON060288, ON060290, ON060291, ON060292, ON060299, ON070301, ON070308.