Z-0288-2008

Abbott Clinical Chemistry Aeroset/Architect (c Systems)Bilirubin Calibrator, List Number (LN) 1E66-03 & List...

Abbott Medical / initiated 2007-08-10 / Terminated / root cause: manufacturing process

Abbott Laboratories Inc. began a recall of Abbott Clinical Chemistry Aeroset/Architect (c Systems)Bilirubin Calibrator, List Number (LN) 1E66-03 & List Number (LN) 1E66-04, Abbott Laboratories, Abbott... on . The reason FDA records is "Incorrect Calibrator values: changes in a raw material used in the Bilirubin Calibrator value assignment process contributed to variability in the preparation of the standard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0288-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott Clinical Chemistry Aeroset/Architect (c Systems)Bilirubin Calibrator, List Number (LN) 1E66-03 & List Number (LN) 1E66-04, Abbott Laboratories, Abbott Park, IL 60064
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
Incorrect Calibrator values: changes in a raw material used in the Bilirubin Calibrator value assignment process contributed to variability in the preparation of the standard.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
An August 10, 2007, Product Correction letter was provided to all ARCHITECT cSystem and AEROSET System customers. The letter advises that changes in a raw material used in the Bilirubin Calibrator value assignment process contributed to variability in the preparation of the NIST standard 916a. Revised Bilirubin values for the calibrator lot numbers in this recall are attached to the recall notice. Six necessary actions were listed in the recall notice as: 1) Identify the lot numbers of Clinical Chemistry Bilirubin Calibrator(s) currently in use. 2) Discard the value sheet and replace with the appropriate values provided with the recall letter. 3) Enter the revised Bilirubin values. 4) Calibrate the assays. 5) Evaluate your quality control. 6) Evaluate your reference range and any shifts in patient results.
Quantity in commerce
7,033 units total worldwide
Distribution
Worldwide: USA, Canada, Mexico, Guatemala, Puerto Rico, Germany, Hong Kong, Singapore, Australia, New Zealand, Jamaica, Venezuela, Columbia, Ecuador, Chile, Uruguay, Argentina, Brazil, Thailand, Trinidad & Tobago, Bahamas, Panama, Cayman Islands, and El Salvador
Product codes and lots
List Number (LN) 1E66-03 Lot Number: 38436M100 (Expires September 30, 2007); List Number (LN) 1E66-04 Lot Numbers: 41456M100 (Expires December 31, 2007), 42396M100 (Expires April 30, 2008), 48616M100 (Expires July 31, 2007)