Z-0070-2008

Medtronic Sprint Fidelis 6949 Steroid eluting, quadripolar, screw-in, ventricular lead with right ventricular...

Medtronic, Inc. / initiated 2007-10-15 / Terminated / root cause: design

Medtronic Inc. Cardiac Rhythm Managment began a recall of Medtronic Sprint Fidelis 6949 Steroid eluting, quadripolar, screw-in, ventricular lead with right ventricular (RV) / Superior Vena Cava (SVC) defibrillation... on . The reason FDA records is "Medtronic Cardiac Rhythm Disease Management has voluntarily suspended worldwide distribution of the Sprint Fidelis family of defibrillation leads because of the potential for lead fractures. In addition, the company...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0070-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic Sprint Fidelis 6949 Steroid eluting, quadripolar, screw-in, ventricular lead with right ventricular (RV) / Superior Vena Cava (SVC) defibrillation coil electrodes. Sterile. This lead is designed for pacing, sensing, cardioversion and defibrillation therapies. Medtronic World Headquarters, Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Medtronic Cardiac Rhythm Disease Management has voluntarily suspended worldwide distribution of the Sprint Fidelis family of defibrillation leads because of the potential for lead fractures. In addition, the company recommends against new implants of the leads (Sprint Fidelis Models: 6930, 6931, 6948, 6949). Medtronic, its Independent Physician Quality Panel, and Bruce Lindsay, M.D., Professor of
Root cause tag
design
FDA root cause
Device Design
Action
A press release was issued by Medtronic on 10/15/2007. An FDA Statement was also issued on 10/15/2007. An Urgent Medical Device Information letter, dated 10/15/2007, was sent to physicians. The letter states that Medtronic has suspended distribution of the leads, the leads should no longer be implanted and unused leads should be returned to Medtronic. This letter describes background information, performance and recommendations. Medtronic believes it is inappropriate to prophylactically remove Sprint Fidelis leads except in unusual individual patient circumstances. An Important Patient Information letter was addressed to Medtronic Heart Device Patients. Both letters can be found on Medtronic's website at www.medtronic.com/fidelis. Return of product will be achieved with the assistance of Medtronic Representatives.
Quantity in commerce
235,000 for all models
Distribution
Worldwide, including USA, Puerto Rico, Guam, and Northern Mariana Islands.
Product codes and lots
all codes