Z-0934-2008

Inotech brand Heparin Adsorbant, Catalog #: HA-1, in 10mm x 75mm polystyrene tubes, each containing 70mg of...

Inotech Biosystems Intl., Inc / initiated 2007-08-20 / Terminated / root cause: labeling and UDI

Inotech Biosystems Intl., Inc began a recall of Inotech brand Heparin Adsorbant, Catalog #: HA-1, in 10mm x 75mm polystyrene tubes, each containing 70mg of Heparin Adsorbent, 10 tubes per foil lined pouch... on . The reason FDA records is "Erroneous Coagulation Tests". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0934-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Inotech Biosystems Intl., Inc
Product
Inotech brand Heparin Adsorbant, Catalog #: HA-1, in 10mm x 75mm polystyrene tubes, each containing 70mg of Heparin Adsorbent, 10 tubes per foil lined pouch, 10 foil pouches/box, Inotech Biosystems International, Inc., Rockville, MD 20855
Product code
JPA System, Multipurpose For In Vitro Coagulation Studies
Reason for recall
Erroneous Coagulation Tests
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
The recalling firm notified consignees by phone on 08/20/07 and provided instructions for increasing the speed of centrifugation from 1,500 x g to 2,500 x g and decanting the supernatant aliquot carefully so as to not draw in resin fibers that may be stuck on the test tube walls during sample preparation for testing in the coagulation studies. The firm updated their recall notification on 10/01/07 as an Urgent Recall/Correction to alert healthcare practitioners to not use "lupus sensitive reagents" with the heparin absorbant product and that Platelet Factor 3 and any other negatively charged molecule in the plasma may interact with the resin causing false high readings due to the partial removal of the clotting initiating factors.
Quantity in commerce
37,370 tubes
Distribution
Worldwide; USA to Hospitals and Clinical Laboratories and to one domestic distributor for export to Canada.
Product codes and lots
Lots #: 120706 and 300407