Z-0183-2008

GE OEC, Insta Trak 3500 Plus with the Cranial Application for Image Guided Surgery, GE Healthcare, Surgery...

GE Healthcare / initiated 2007-10-18 / Terminated / root cause: design

GE OEC Medical Systems, Inc began a recall of GE OEC, Insta Trak 3500 Plus with the Cranial Application for Image Guided Surgery, GE Healthcare, Surgery, Salt Lake City, UT 84116 on . The reason FDA records is "Crack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0183-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE OEC, Insta Trak 3500 Plus with the Cranial Application for Image Guided Surgery, GE Healthcare, Surgery, Salt Lake City, UT 84116
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Crack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were sent an Urgent Recall Notice via mail dated 10/11/07, notifying the customer to examine the plastic end of the transmitter arm and to not use the transmitter arm, if any cracks or damage are observed. Also to stop using Steam Sterilization method and to utilize the other Sterilization methods indicated in the Operators Manual. The root cause is still under investigation GEHC Surgery will provide, at no charge, a solution to this issue when it becomes available.
Quantity in commerce
169 units
Distribution
Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IL, MA, MD, MI, MS, NC, NY, OK, OR, PA, SC, TX, UT, VA, WI, and WV, and countries of Canada, China, France, Hong Kong, Japan, Mexico, Spain, and Switzerland.
Product codes and lots
Not applicable since this is an arm for all devices listed in product description.