Z-0617-2008
Hill-Rom Affinity 4 Birthing Bed; Model P3700B, Hill-Rom, Inc., Batesville, IN 47006
Hill-Rom Inc. / initiated 2007-09-06 / Terminated / root cause: design
Hill-Rom, Inc. began a recall of Hill-Rom Affinity 4 Birthing Bed; Model P3700B, Hill-Rom, Inc., Batesville, IN 47006 on . The reason FDA records is "Brake failure: The brakes may not hold or lock properly". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0617-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Hill-Rom Affinity 4 Birthing Bed; Model P3700B, Hill-Rom, Inc., Batesville, IN 47006
- Product code
- HDD Table, Obstetrical, Ac-Powered (And Accessories)
- Reason for recall
- Brake failure: The brakes may not hold or lock properly.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Hill-Rom notified consignees via Urgent Medical Device Correction letter dated 9/6/07. The letter instructed facilities to continue using the beds if they pass a test of the brake mechanism; and a Hill-Rom representative will contact the facility to schedule a replacement of the detent mechanism and adjustment of the casters.
- Quantity in commerce
- 1,796
- Distribution
- Worldwide Distribution - USA, Argentina, Australia, Bahamas, Belgium, Belize, Brazil, Canada, China, Denmark, Dominican Republic, France, Germany, Guam, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jordan, Kenya, Malaysia, Mexico, Netherlands, Norway, Philippines, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Syria, Turkey, United Arab Emirates and United Kingdom.
- Product codes and lots
- All units manufactured prior to 11/20/07; all beds with serial numbers below 1232AA6856.