Z-0446-2008

Dale 240 Blue (Adults) Tracheostomy Tube Holder, Product Number: H84102401, Dale Medical Products, Inc...

Dale Medical Products, Inc. / initiated 2007-09-27 / Terminated / root cause: labeling and UDI

Dale Medical Products, Inc. began a recall of Dale 240 Blue (Adults) Tracheostomy Tube Holder, Product Number: H84102401, Dale Medical Products, Inc., Plainville, MA 02762 on . The reason FDA records is "Revised Labeling : Instructions For use: "Do not trim fastener hook tabs. " Trimming can cause tabs to disengage at the trach plate". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0446-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Dale Medical Products, Inc.
Product
Dale 240 Blue (Adults) Tracheostomy Tube Holder, Product Number: H84102401, Dale Medical Products, Inc., Plainville, MA 02762
Product code
CBH Device, Fixation, Tracheal Tube
Reason for recall
Revised Labeling : Instructions For use: "Do not trim fastener hook tabs. " Trimming can cause tabs to disengage at the trach plate
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Dale Medical notified Distributors by letter dated 9/27/07, identified as an Urgent Medical Device Correction Letter. The letter requested distributors to notify all their customers by letter. The Correction Letter reminded users to take notice of new labeling " Do not trim fasterner hook tabs"
Quantity in commerce
600,000 units
Distribution
Worldwide-USA and countries of Canada, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Iceland, Italy, The Netherlands, Norway, The UK, Spain, Switzerland, Australia, Hong Kong,India, Japan, Korea, Malaysia, New Zealand, The Philippines,Singapore, Taiwan, ROC, Brazil, Bermuda, Chile, Costa Rica,Ecuador, Mexico, Panama, Venezuela, Thailand, Saudi Arabia, and The United Arab Emirates.
Product codes and lots
All lots