Z-0434-2008

Depuy TK2 Compression Hip Screw trauma plate, Standard Barrel,135 Degree, 4H-92.6mm, sterile; REF...

Depuy Orthopaedics Inc. / initiated 2007-10-30 / Terminated / root cause: manufacturing process

Depuy Orthopaedics, Inc. began a recall of Depuy TK2 Compression Hip Screw trauma plate, Standard Barrel,135 Degree, 4H-92.6mm, sterile; REF 8315-35-004, DePuy Orthopaedics, Inc., Warsaw, IN 46581 on . The reason FDA records is "Mis-etched/labeled as to barrel length: the lot with the short barrel was etched and labeled as a standard barrel and the lot with the standard barrel was etched and labeled as a short barrel". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0434-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Depuy TK2 Compression Hip Screw trauma plate, Standard Barrel,135 Degree, 4H-92.6mm, sterile; REF 8315-35-004, DePuy Orthopaedics, Inc., Warsaw, IN 46581
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
Mis-etched/labeled as to barrel length: the lot with the short barrel was etched and labeled as a standard barrel and the lot with the standard barrel was etched and labeled as a short barrel.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified via Urgent Information, Recall Notice(letter) dated 10/30/07 to determine whether they have product on hand and to contact their sales representative to return the product. Dear Doctor letters will be issued to implanting physicians as they are identified.
Quantity in commerce
9
Distribution
Nationwide including the states of Florida, Massachusetts, Ohio, Pennsylvania, South Carolina, Tennessee and Virginia.
Product codes and lots
Lot DHDB73.