Hypoguard Advance Micro-draw Blood Glucose Monitoring System, Product Number : 411100, For In Vitro...
ARKRAY USA INC. began a recall of Hypoguard Advance Micro-draw Blood Glucose Monitoring System, Product Number : 411100, For In Vitro Diagnostic Use, Manufactured for ARKRAY USA, Inc... on . The reason FDA records is "Mis-labeled/coded strips : Meter kits from this lot contain mislabeled 10-count test strip vials. The code number printed on the Advance Micro-draw 10-count test strip bottle is incorrect. The code number on the vial is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0458-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arkray Factory USA, Inc.
- Product
- Hypoguard Advance Micro-draw Blood Glucose Monitoring System, Product Number : 411100, For In Vitro Diagnostic Use, Manufactured for ARKRAY USA, Inc., Minneapolis, MN 55439 USA. (Package Contents -Advance Micro-draw Blood Glucose Meter, 10 test Strips, Codeing chip and Control Solution (3mL), Lancing device, 10 Lancets, 3V Lithium Battery, Logbook, User Instruction Manual, Quick Reference Guide)
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- Mis-labeled/coded strips : Meter kits from this lot contain mislabeled 10-count test strip vials. The code number printed on the Advance Micro-draw 10-count test strip bottle is incorrect. The code number on the vial is 867 when it should be 687. The 687-labeled code chip contains the correct calibration information for the "867" test strips, so the meter will function properly and give accurat
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Distributors were sent a letter on 10/18/07 asking that product be quarantined. On 10/26/07, an Urgent Voluntary Medical Device Recall letter was sent asking that they return affected product and contact their customers to the retail level.
- Quantity in commerce
- 2,005 Kits
- Distribution
- Nationwide: USA including states of CA, MI, MO, MS, NE, OH and TX.
- Product codes and lots
- Part Number 411100, Lot Number: 09177A