VKS Tibial Trial, Right/Left, Prep Guide; Model Number: 1214331, 1214332, 1214333, 1214334, 1214335: ( Model...

Plus Orthopedics / initiated 2006-11-27 / Terminated / root cause: manufacturing process

Plus Orthopedics USA began a recall of VKS Tibial Trial, Right/Left, Prep Guide; Model Number: 1214331, 1214332, 1214333, 1214334, 1214335: ( Model Number: 1214331 VKS Tibial Prep Guide, Right/Left... on . The reason FDA records is "The VKS "Tibial Trial" was found to have an incorrect fixation pin hole diameter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0362-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Plus Orthopedics
Product
VKS Tibial Trial, Right/Left, Prep Guide; Model Number: 1214331, 1214332, 1214333, 1214334, 1214335: ( Model Number: 1214331 VKS Tibial Prep Guide, Right/Left, Size 1, Lot Code EM. 1 and EM.6; 1214332 VKS Tibial Prep Guide, Right/Left, Size 2 Lot Code EM.l and EM.6, 1214333 VKS Tibial Prep Guide, Right/Left, Size 3 Lot Code EM.1; 1214334 VKS Tibial Prep Guide, Right/Left, Size 4 Lot Code EM.5; 1214335 YKS Tibial Prep Guide, Right/Left, Size 5 Lot Code EM), Plus Orthopedics
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The VKS "Tibial Trial" was found to have an incorrect fixation pin hole diameter.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The initial depth of the withdrawal action taken November 2006 was to the distributor (agent) level. On 11/27/2006 Plus Orthopedics USA contacted individual agent sites showing inventory of the affected article numbers on hand and requested that those items be returned to Plus USA. Agents were notified by telephone to check their inventory and return any affected units. A telephone script was used.
Quantity in commerce
19
Distribution
Nationwide to : AZ, CA, CO, FL, ID, NE, NY, SC, TX, UT, VA, and WA
Product codes and lots
Model Numbers/Lot Codes: 1214331/EM.1 and EM.6, 1214332/EM.1 and EM.6, 1214333/EM.1, 1214334/EM.5, and 1214335/EM.1