Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton...
Medtronic Navigation, Inc began a recall of Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0) on . The reason FDA records is "Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0294-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0)
- Product code
- JZL
- Reason for recall
- Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Medtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.
- Quantity in commerce
- 7 units
- Distribution
- Worldwide: USA including states of CO, IN, NE, and MN and countries of Canada and Germany
- Product codes and lots
- System Serial Numbers: 103, 104, 109, 117, 119 (Germany) 125, and 144 (Canada)