Z-0294-2008

Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton...

Medtronic Navigation, Inc / initiated 2007-10-17 / Terminated / root cause: software

Medtronic Navigation, Inc began a recall of Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0) on . The reason FDA records is "Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0294-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0)
Product code
JZL
Reason for recall
Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)
Root cause tag
software
FDA root cause
Software change control
Action
Medtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.
Quantity in commerce
7 units
Distribution
Worldwide: USA including states of CO, IN, NE, and MN and countries of Canada and Germany
Product codes and lots
System Serial Numbers: 103, 104, 109, 117, 119 (Germany) 125, and 144 (Canada)