Si emans Simview NT Cassette Holder Electronic Imaging Device (Image Identifier) with IDI, Part No: 54 97...
Siemens Medical Solutions USA, Inc began a recall of Si emans Simview NT Cassette Holder Electronic Imaging Device (Image Identifier) with IDI, Part No: 54 97 016; Siemens Medical Solutions USA, Concord, CA 94520 on . The reason FDA records is "Loose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0284-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Si emans Simview NT Cassette Holder Electronic Imaging Device (Image Identifier) with IDI, Part No: 54 97 016; Siemens Medical Solutions USA, Concord, CA 94520
- Product code
- KPQ System, Simulation, Radiation Therapy
- Reason for recall
- Loose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Customers were sent a Customer Safety Advisory Notice on February 27, 2007, notifying them that the mounting screws on the cassette holders on the SIMVIEW 3000 and NT Simulators, as well as some MEVASIM Simulators, could loosen and can cause damage to the system, or personal injury. The firm requested that they check of tightness on the mounting hardware. A STOP Modification notice was issued on 06/01/2007, to all affected regional service organizations that are responsible for implementing updates due to new information. Updated letters and User instructions will be released in August 2007.
- Quantity in commerce
- 287 units total, all recalls Z-0281-285-2008
- Distribution
- Worldwide Distribution - USA including states of IN, VA, MA, VT, NJ, NY, PA, DC, NC, SC, GA, AL, TN, MI, KY, OH, IN MI, MN, ND, MT, IL, MO, KS, NE, AR, OK, TX, AZ, CA, SD, LA, Puerto Rico, and countries of Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Colombia, Costa Rica, Croatia, Denmark, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Hungary, India, Iran, Ireland, Italy, Kuwait, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, China, Israel/Palestine, Philippines, Poland, Portugal, Qatar, Republic of Korea, Russia, Singapore, South Africa, Spain, Thailand, Turkey, Ukraine, UK, and Vietnam.
- Product codes and lots
- Serial Numbers: 6018, 6001, 6008, 6067, 6037, 6038, 6059, 6041, 6047, 6017, 6027, 6036, 6040, 6050, 1256, 6082, 6056, 6064, 6083, 6034, 6044, 6055, 6058, 6028, 6090, 6105, 2255, 6079, 6089, 6073, 6052, 6077, 6072, and 6066.