Z-0287-2008
Ellex Integre Duo, Model No: LP1RG - Ellex Medical PTY, LTD 82 Gilbert Street, Adelaide, S.A. 5000...
Ellex USA / initiated 2007-06-13 / Terminated / root cause: design
Ellex USA began a recall of Ellex Integre Duo, Model No: LP1RG - Ellex Medical PTY, LTD 82 Gilbert Street, Adelaide, S.A. 5000, AUSTRALIA. (The Integre Duo is a dual-wavelength surgical... on . The reason FDA records is "Electric Shock from Power Supply: The power supply has a defect which can cause an electrical shock to user". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0287-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ellex USA
- Product
- Ellex Integre Duo, Model No: LP1RG - Ellex Medical PTY, LTD 82 Gilbert Street, Adelaide, S.A. 5000, AUSTRALIA. (The Integre Duo is a dual-wavelength surgical photocagulator laser intended for use in a clinic, outpatient clinic, or a retinal specialist's office. )
- Product code
- HQF Laser, Ophthalmic
- Reason for recall
- Electric Shock from Power Supply: The power supply has a defect which can cause an electrical shock to user.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Ellex, Inc. called each customer on 6/13/07, about the problem (defect which can cause electric shock) and advised that they would be performing modification of the power supply in the laser system and scheduling a field service visit to complete the work. Upon completion of the modification, the service technician completed the service record and left a copy with each client and informed them of the completed work. A copy of each completed service record was sent to Ellex Medical Pty. Ltd. On 9/27/07, Ellex sent a letter to all consignees, informing them that all modifications by service technicians' have been completed, and that their Solitaire LP4532 laser and/or Intregre Duo LP1RG is up to the manufacturers specifications and is in good operating order.
- Quantity in commerce
- 1
- Distribution
- Worldwide Distribution - USA including states of WI, PA, CO, CA, NC, NY, SC, MD, and MS, and countries of Canada and Brazil.
- Product codes and lots
- Integre Duo LP1RG-S P/N 3040000 (serial # /system #): 31418 / ID1057