Z-0478-2008
Hallmark Anterior Cervical Plate System: Primary Bone Screw, Semi Constrained, REF # 62-3140,14mm; Blackstone...
Blackstone Medical, Inc. / initiated 2006-05-08 / Terminated / root cause: design
Blackstone Medical, Inc. began a recall of Hallmark Anterior Cervical Plate System: Primary Bone Screw, Semi Constrained, REF # 62-3140,14mm; Blackstone Medical, 90 Brookdale Drive, Springfield, MA... on . The reason FDA records is "Incomplete Design Validation: the laser marking on the head of the screw lacked complete process qualification". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0478-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Blackstone Medical, Inc.
- Product
- Hallmark Anterior Cervical Plate System: Primary Bone Screw, Semi Constrained, REF # 62-3140,14mm; Blackstone Medical, 90 Brookdale Drive, Springfield, MA 01104 (The Hallmark Anterior Cervical Plate Systems consists of an assortment of implantable titanium alloy plates and bone screws that are provided non-sterile)
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- Incomplete Design Validation: the laser marking on the head of the screw lacked complete process qualification
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On May 8, 2006, a communication to the field was sent via phone, e-mail and fax. Instruction was made to retrieve and return all product in the field.
- Quantity in commerce
- 1,991 semi-constrained primary and rescue screws.
- Distribution
- Nationwide
- Product codes and lots
- All Lots (Primary semi-contained rescue Screws: P/N's 62-3100 through 62-3180 and rescue simi-constrained bone screws P/N's 62-4100- 62-4180)