Z-0554-2008

Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models...

Varian Medical Systems, Inc. / initiated 2007-09-26 / Terminated / root cause: software

Varian Medical Systems Inc began a recall of Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models; Varian Medical Systems, Palo Alto, CA 94304 on . The reason FDA records is "Erroneously Resetting: RPM System version 1.7.3, software failure can occur that affects the gated radiotherapy treatment when phase-based gating is used. RPM 1.7 erroneously resets the phase-based gating thresholds to...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0554-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models; Varian Medical Systems, Palo Alto, CA 94304
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Erroneously Resetting: RPM System version 1.7.3, software failure can occur that affects the gated radiotherapy treatment when phase-based gating is used. RPM 1.7 erroneously resets the phase-based gating thresholds to system default during the treatment session, if the user accesses the Session options during the treatment session. This can lead to difference between the planned and delivered
Root cause tag
software
FDA root cause
Software design
Action
Varian sent an Urgent Medical Device Correction letter on September 26, 2007, to all its direct consignees, informing them that the RPM version 1.7 may reset the phase thresholds if Sessions options are accessed during treatment. A Service Representative will be scheduling a visit to user site to perform the modification requested to address the problem, when a corrected software version is available.
Quantity in commerce
59 units
Distribution
Worldwide including countries of: Ireland, Italy, Spain, UK, Denmark, Finland, France, and Canada
Product codes and lots
Serial Number:  H502244, H502211, H502203, H502183, H502179, H502178, H502169, H502167, H502166, H502154, H502152, H502138, H502137, H502136, H502133, H502132, H502131, H502130, H502126, H502121, H502119, H502118, H502117, H502116, H502112, H502110, H502107, H502104, H502103, H502102, H502096, H502060, H502057, H502045, H502002, H501999, H501983, H501930, H501907, H501904, H501869, H501838, H501577, H501412, H501367, H501329, H501325, H501306, H501290, H501259, H501246, H501222, H501214, H502051, H502067, H500395, H500406, H500853, and H502066