Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models...
Varian Medical Systems Inc began a recall of Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models; Varian Medical Systems, Palo Alto, CA 94304 on . The reason FDA records is "Erroneously Resetting: RPM System version 1.7.3, software failure can occur that affects the gated radiotherapy treatment when phase-based gating is used. RPM 1.7 erroneously resets the phase-based gating thresholds to...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0554-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian Real-Time Position Management System (RPM), A Respiratory Gating System; Version 1.7.3, All Models; Varian Medical Systems, Palo Alto, CA 94304
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Erroneously Resetting: RPM System version 1.7.3, software failure can occur that affects the gated radiotherapy treatment when phase-based gating is used. RPM 1.7 erroneously resets the phase-based gating thresholds to system default during the treatment session, if the user accesses the Session options during the treatment session. This can lead to difference between the planned and delivered
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Varian sent an Urgent Medical Device Correction letter on September 26, 2007, to all its direct consignees, informing them that the RPM version 1.7 may reset the phase thresholds if Sessions options are accessed during treatment. A Service Representative will be scheduling a visit to user site to perform the modification requested to address the problem, when a corrected software version is available.
- Quantity in commerce
- 59 units
- Distribution
- Worldwide including countries of: Ireland, Italy, Spain, UK, Denmark, Finland, France, and Canada
- Product codes and lots
- Serial Number: H502244, H502211, H502203, H502183, H502179, H502178, H502169, H502167, H502166, H502154, H502152, H502138, H502137, H502136, H502133, H502132, H502131, H502130, H502126, H502121, H502119, H502118, H502117, H502116, H502112, H502110, H502107, H502104, H502103, H502102, H502096, H502060, H502057, H502045, H502002, H501999, H501983, H501930, H501907, H501904, H501869, H501838, H501577, H501412, H501367, H501329, H501325, H501306, H501290, H501259, H501246, H501222, H501214, H502051, H502067, H500395, H500406, H500853, and H502066