Varian brand Clinac, Accelerator, Linear, Medical charged-particle radiation therapy system; Model: Low...
Varian Medical Systems Inc began a recall of Varian brand Clinac, Accelerator, Linear, Medical charged-particle radiation therapy system; Model: Low Energy Clinacs with one or more of the following... on . The reason FDA records is "The collimator drive chain may break or slip off of its drive track, allowing the collimator to rotate freely without motor control; if undetected resulting in a treatment with the wrong collimator angle". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0545-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian brand Clinac, Accelerator, Linear, Medical charged-particle radiation therapy system; Model: Low Energy Clinacs with one or more of the following options: -- BrainLAB micro MLC -- Stereotactic motion disable Product is manufactured and distributed by Varian Medical Systems, Palo Alto, CA
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- The collimator drive chain may break or slip off of its drive track, allowing the collimator to rotate freely without motor control; if undetected resulting in a treatment with the wrong collimator angle.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- On 9/10/07, the firm issued an Urgent Medical Device Correction letter to all its direct consignees, informing them of the device defect and providing instructions on the recall. Short term user instructions were provided and the firm will schedule follow-up visits to make the needed permanent device corrections at each device site.
- Quantity in commerce
- 418 (total for Recalls Z-0544-0545-2008)
- Distribution
- Worldwide- Canada, Mexico, and the USA
- Product codes and lots
- The Serial Numbers: H180001, H180007, H180076, H180143, H180147, H180169, H180181, H180187, H180195, H180222, H180273, H180276, H180311, H180314, H180322, H180460, H180467, H180473, H180490, H180493, H180514, H180533, H180550, H180563, H180566, H180570, H180576, H180599, H180613, H180631, H180671, H180690, H180697, H180718, H180731, H180739, H180779, H180785, H180786, H180808, H180815, H180816, H180822, H180836, H180844, H180855, H180877, H180894, H181065, H180066, and H180769