Ascent POCTS (Posterior Occipital Cervical Thoracic System) Part Numbers: 65-33 XX Series (65-3310,-3312...
Blackstone Medical, Inc. began a recall of Ascent POCTS (Posterior Occipital Cervical Thoracic System) Part Numbers: 65-33 XX Series (65-3310,-3312, -3314, -3318, -3320-, -3322, -3324, -3326, -3328, and... on . The reason FDA records is "Manufacturing Defect: The saddle of the screw's body relative to the keyway exceeds the tolerance outlined in the device specifications. The extent to which the tolerance is exceeded varies from part to part, in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0363-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Blackstone Medical, Inc.
- Product
- Ascent POCTS (Posterior Occipital Cervical Thoracic System) Part Numbers: 65-33 XX Series (65-3310,-3312, -3314, -3318, -3320-, -3322, -3324, -3326, -3328, and -3330), Multi-Axial Screw, 4.0mm x 26mm;Implant grade Titanium Alloy; Blacksone Medical Inc., 90 Brookdale Drive, Springfield, MA 01104.
- Product code
- MNI Orthosis, Spinal Pedicle Fixation
- Reason for recall
- Manufacturing Defect: The saddle of the screw's body relative to the keyway exceeds the tolerance outlined in the device specifications. The extent to which the tolerance is exceeded varies from part to part, in the extreme the tolerance is exceeded by (+) 50%. As a consequence, there is a potential for set-screw cross-threading during implantation. Cross-threading could in turn lead to a comprom
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Blackstone Medical, Inc., field representatives were notified via phone, e-mail or fax on 4/30/07. Attention: Notification of Product Removal Letters were sent on May 15, 2007, to distributors and implanting surgeons requesting return of non-implanted stock.
- Quantity in commerce
- 319 screws ( 212 domestic, 107 international)
- Distribution
- Worldwide: USA and countries of China, Cyprus, Turkey and Switzerland.
- Product codes and lots
- Part Numbers: 65-3310,-3312, -3314, -3318, -3320-, -3322, -3324, -3326, -3328, and -3330; Lot numbers: 009, 010, 011, 018, 019, 031, 048, 064, 084 and 107.
FIRST ARTICLE / WEEKLY