Z-0296-2008

Xoran MiniCAT for ENT, Xoran Technologies, Incorporated, Ann Arbor, MI 48103. This device is a cone beam CT...

Xoran Technologies, LLC / initiated 2006-07-13 / Terminated

Xoran Technologies, Inc. began a recall of Xoran MiniCAT for ENT, Xoran Technologies, Incorporated, Ann Arbor, MI 48103. This device is a cone beam CT scan for sinue, temporal bones and maxillofacial... on . The reason FDA records is "The hard drive may crash, resulting in loss of data from patient CT scans". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0296-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Xoran Technologies, LLC
Product
Xoran MiniCAT for ENT, Xoran Technologies, Incorporated, Ann Arbor, MI 48103. This device is a cone beam CT scan for sinue, temporal bones and maxillofacial imaging.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The hard drive may crash, resulting in loss of data from patient CT scans.
FDA root cause
Finished device change control
Action
Consignee accounts were visited between 7/25/06 and 8/10/07 and the hard drives on their systems were replaced.
Quantity in commerce
5
Distribution
Nationwide Distribution: California, South Dakota, Tennessee, Utah and West Virginia.
Product codes and lots
Serial Numbers: ENT00009, ENT00051, ENT00084, ENT00086 and ENT00088.