Triathlon MIS Modular Capture;Catalog Number: 6541-5-723; Stryker Orthopaedics, Mahwah, New Jersey
Stryker Howmedica Osteonics Corp. began a recall of Triathlon MIS Modular Capture;Catalog Number: 6541-5-723; Stryker Orthopaedics, Mahwah, New Jersey on . The reason FDA records is "Assembly problems: MIS instruments (Triathlon MIS AP Sizer Body - Left; Triathlon MIS Sizer Body - Right; Triathlon MIS Femoral Adjustment Block and Triathlon MIS Modular Capture) may not assemble properly". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0450-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Triathlon MIS Modular Capture;Catalog Number: 6541-5-723; Stryker Orthopaedics, Mahwah, New Jersey
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Assembly problems: MIS instruments (Triathlon MIS AP Sizer Body - Left; Triathlon MIS Sizer Body - Right; Triathlon MIS Femoral Adjustment Block and Triathlon MIS Modular Capture) may not assemble properly.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Important Product Correction letters and Product Accountability Forms were sent via Federal Express on 9/19/06 with return receipt. Foreign locations were notified by a Advisory Notice. Notification letters requested that all inventory of the affected items be examined to verify that it assemble correctly. Any product that does not assemble correctly was to be returned to Stryker, Mahwah, NJ. The Product Accountability form was to be completed by each location to notify Stryker that the letter ahd been received and inventory examined. Stryker representatives were to be contacted for product replacement.
- Quantity in commerce
- 2,868 units total for Recalls Z-0448-0450-2008
- Distribution
- Worldwide: USA including states of MI and IL. and countries of Australia, China, New Zealand, Singapore, France, Germany, Italy, Portugal, Spain, Sweden and UK.
- Product codes and lots
- Catalog Number: 6541-5-723; All lot codes.