bioMerieux, BacT/ALERT FA Culture Bottles, bioMerieux, Inc., Durham, NC 27704
bioMerieux, Inc. began a recall of bioMerieux, BacT/ALERT FA Culture Bottles, bioMerieux, Inc., Durham, NC 27704 on . The reason FDA records is "Duplicate Bottle Indentification -- Three lot numbers of BacT/ALERT FA Culture Bottles labels contained duplicate bottle identification (Bottle ID) barcode numbers. (Lots 1017685 and 1017788 contain the same series of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0457-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- bioMerieux, BacT/ALERT FA Culture Bottles, bioMerieux, Inc., Durham, NC 27704
- Product code
- MDB System, Blood Culturing
- Reason for recall
- Duplicate Bottle Indentification -- Three lot numbers of BacT/ALERT FA Culture Bottles labels contained duplicate bottle identification (Bottle ID) barcode numbers. (Lots 1017685 and 1017788 contain the same series of Bottle ID's; Lot 1017693 contains duplicate Bottle ID's within the lot and may share Bottle ID's with both lots 1017685 and 1017788)
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- bioMerieux notified all Consignees with a Urgent Product Recall Notice letter on 11/05/2007. They were instructed to examine their inventory to determine if they have any remaining inventory of the 3 affected lots. If they have any of these bottles they were instructed to stop using them and destroy all remaining inventory. A Customer Acknowledgement Form was attached to be completed and returned to bioMerieux via fax or mail. There were also instructions attached for the use of Generic Bottle Barcodes and Creating a Unique Bottle ID.
- Quantity in commerce
- 7,025
- Distribution
- Worldwide Distribution-USA including countries of Australia, Austria, Bangladesh, Brazil, Bulgaria, Byelorussia, Chile, China, Colombia, Croatia, Denmark, Dijbouti, Estonia, Federation of Russia, Finland, France, Germany, Greece, Guadeloupe, India, Israel, Italy, Korea, Kuwait, Martinique, Mexico, Netherlands, New Caledonia, Norway, Panama, Poland, Portugal, Reunion, Saudi Arabia, Serbia Montenegro, Slovenia, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Vietnam.
- Product codes and lots
- Lot Numbers: 1017685, 101793 and 1017788