Z-0377-2008

Roche Ammonia/Ethanol/CO2 Calibrator for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and...

Roche Diagnostics GmbH / initiated 2007-11-14 / Terminated / root cause: manufacturing process

Roche Diagnostics Corp. began a recall of Roche Ammonia/Ethanol/CO2 Calibrator for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas c 501 clinical chemistry analyzers; Catalog... on . The reason FDA records is "High Quality Control Recovery: Undetected high quality control recovery in bicarbonate assay". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0377-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Ammonia/Ethanol/CO2 Calibrator for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas c 501 clinical chemistry analyzers; Catalog No. 20751995190. Roche Diagnostics Corporation, Indianapolis, IN 46250
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
High Quality Control Recovery: Undetected high quality control recovery in bicarbonate assay.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Consignees were notified of this recall by an Urgent Medical Device Recall letter sent on 11/14/07. The letter instructed the users to immediately discontinue use of Ammonia/ Ethanol/CO2 Control N and Ammonia/Ethanol/CO2 Control A (with listed lots) for quality control assessment of the bicarbonate assay. They were informed that controls may still be used to assess ammonia and ethanol, but alternative controls must be used for assessment of the bicarbonate assay. Consignees are instructed to destroy products at their facility. Consignees were notified of the availability of replacement product via letter dated .
Quantity in commerce
1,904
Distribution
Nationwide Distibution
Product codes and lots
Lot No: 179070-01 and 179070-02; exp. 4/30/2008.