Z-0378-2008
Roche Ammonia/Ethanol/CO2 Control N (Normal) for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas...
Roche Diagnostics GmbH / initiated 2007-11-14 / Terminated / root cause: manufacturing process
Roche Diagnostics Corp. began a recall of Roche Ammonia/Ethanol/CO2 Control N (Normal) for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas c 501 clinical chemistry analyzers... on . The reason FDA records is "High Quality Control Recovery: Undetected high quality control recovery in bicarbonate assay". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0378-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche Ammonia/Ethanol/CO2 Control N (Normal) for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas c 501 clinical chemistry analyzers; Catalog. No. 20752401190. Roche Diagnostics Corporation, Indianapolis, IN 46250
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- High Quality Control Recovery: Undetected high quality control recovery in bicarbonate assay.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Consignees were notified of this recall by an Urgent Medical Device Recall letter sent on 11/14/07. The letter instructed the users to immediately discontinue use of Ammonia/ Ethanol/CO2 Control N and Ammonia/Ethanol/CO2 Control A (with listed lots) for quality control assessment of the bicarbonate assay. They were informed that controls may still be used to assess ammonia and ethanol, but alternative controls must be used for assessment of the bicarbonate assay. Consignees are instructed to destroy products at their facility. Consignees were notified of the availability of replacement product via letter dated .
- Quantity in commerce
- 811
- Distribution
- Nationwide Distibution
- Product codes and lots
- Lot No: 178993-01 and 178993-02; exp. 3/31/2008.