Z-0531-2008

Terumo ultrasonic air sensor, 1/4" by 1/16" (6.4 mm x 1.6 mm), Black, for use on Terumo perfusion systems 1...

Terumo Cardiovascular Systems Corporation / initiated 2007-09-19 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corp began a recall of Terumo ultrasonic air sensor, 1/4" by 1/16" (6.4 mm x 1.6 mm), Black, for use on Terumo perfusion systems 1, 8000 and 9000; Catalog No: 5785, Terumo... on . The reason FDA records is "False Alarm: The air sensor may malfunction and trigger false alarms, and may continue to alarm thus preventing the device from being reset". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0531-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo ultrasonic air sensor, 1/4" by 1/16" (6.4 mm x 1.6 mm), Black, for use on Terumo perfusion systems 1, 8000 and 9000; Catalog No: 5785, Terumo Cardiovalvular Systems, Ann Arbor, MI 48103
Product code
KRL Detector, Bubble, Cardiopulmonary Bypass
Reason for recall
False Alarm: The air sensor may malfunction and trigger false alarms, and may continue to alarm thus preventing the device from being reset.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified of the defect/issue and what to do if it were to occur (short term work around) via an Urgent Medical Device Removal letter dated 9/14/07. As the permanent fix users were informed the sensor will be replaced when replacements become available.
Quantity in commerce
232
Distribution
Worldwide: USA, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, Costa Rica, Dominican Republic, Equador, Germany, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, United Arab Emirates, Uruguay, Venezuela and Vietnam.
Product codes and lots
Serial Numbers: 1106 through 1370 and 1376.