Hill-Rom Affinity Three Birthing Bed, Model P3700A

Hill-Rom Inc. / initiated 2007-11-21 / Terminated / root cause: design

Hill-Rom, Inc. began a recall of Hill-Rom Affinity Three Birthing Bed, Model P3700A on . The reason FDA records is "Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0583-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
Hill-Rom Affinity Three Birthing Bed, Model P3700A.
Product code
KNC Table, Obstetric (And Accessories)
Reason for recall
Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified of the problem via an Urgent Medical Device Correction letter dated 11/21/07. Users were informed that they would be receiving a free upgrade kit for all of their Affinity Birthing Beds. The Kit includes two plugs per bed, an insertion tool, and full written instructions. An instruction video is available at www.Hill-Rom.com.
Quantity in commerce
32,379 for all model numbers
Distribution
Worldwide Distribution-- including USA and countries of Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guam, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Morocco, The Netherlands, New Calidonia, New Zealand, Nigeria, Northern Mariana Islands, Norway, Oman, Panama, Philippines, Poland, Portugal, Qatar, Republic of Korea, Saudi Arabia, Singapore, South Korea, Romania, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguey, Venezuela, and Virgin Islands .
Product codes and lots
All units. (**Affinity beds manufactured with serial number I290AA7395 or greater are not impacted by this correction.)