Z-0292-2008
Medrad Avanta Multi-Patient Disposable Hand Controller, Angiographic Injector and Syringe. Catalog number AVA...
Medrad, Inc. / initiated 2007-10-25 / Terminated / root cause: packaging
Medrad Inc began a recall of Medrad Avanta Multi-Patient Disposable Hand Controller, Angiographic Injector and Syringe. Catalog number AVA 500 HC, For Use with MEDRAD Avant Fluid... on . The reason FDA records is "Sterility (package integrity) compromised: Incomplete package seal; sterility may be compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0292-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medrad, Inc.
- Product
- Medrad Avanta Multi-Patient Disposable Hand Controller, Angiographic Injector and Syringe. Catalog number AVA 500 HC, For Use with MEDRAD Avant Fluid Management System, MEDRAD, Inc., Indianola, PA 15051
- Product code
- DXT Injector And Syringe, Angiographic
- Reason for recall
- Sterility (package integrity) compromised: Incomplete package seal; sterility may be compromised
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- The recalling firm issued an Urgent Medical Device Field Correction Letter dated 10/25/07 via Federal Express to their customers informing them of the problem. Customers can either return the product or consistent with the indicated use and established pratice of sheathing, utilize sterile sheaths with the hand controllers for up to five patients. Each letter is being accompanied by a box of 25 sheaths.
- Quantity in commerce
- 25, 977 units
- Distribution
- Worldwide-USA and countries of Germany, Israel, Finalnd, Italy, Netherland, Austria, France, Japan, Denmark, Ireland, South Africia, Australia, Saudi Arabia, Singapore, Canada, New Zealand, Spain, Sweden, Portugal, Korea, Great Britain, Turkey and The United Arab Emerites.
- Product codes and lots
- Catalog Number : AVA500HC. Lot Numbers: 0906AACL01, 0906AACL02, 1006AACL01, 1006AACL02, 1006AACL03, 1106AACL01, 1106AACL03, 1106AACL04, 1106AACL05, 1206AACL01, 1206AACL02, 1206AACL03, 0107AACL01, 0107AACL02, 0107AACL03, 0107AACL04, 0107AACL05, 0107AACL06, 0207AACL01, 0307AACL01, 0307AACL02, 0307AACL03, 0407AACL01, 0407AACL02, 0507AACL01, 0507AACL02, 0607AACL01, 0607AACL02, 0607AACL03, and 0707AACL01.