Z-0115-2008

Biomet Straight Magnum Inserter Handle, a preparatory instrument used for acetabular component insertion; REF...

Zimmer GmbH / initiated 2007-09-04 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Biomet Straight Magnum Inserter Handle, a preparatory instrument used for acetabular component insertion; REF # or Catalog Number: S313141. (systems affected... on . The reason FDA records is "Binding: The instrument was manufactured incorrectly and its use may result in binding to the acetabular component and cause the implant to not release following impaction". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0115-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Straight Magnum Inserter Handle, a preparatory instrument used for acetabular component insertion; REF # or Catalog Number: S313141. (systems affected by this action include M2a-Magnum, Regenerex, Acetabular Revision Shell and M2a-38), Biomet Orthopedics, Inc., Warsaw, IN. 46581
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Binding: The instrument was manufactured incorrectly and its use may result in binding to the acetabular component and cause the implant to not release following impaction.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Consignees were notified by Urgent Medical Device Correction Notice (letter) dated 9/4/07 to discontinue use of the instruments and to return them for evaluation.
Quantity in commerce
22
Distribution
Nationwide; including states of Alabama, Arizona, Arkansas, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Michigan, Nevada, North Carolina, Ohio and South Dakota.
Product codes and lots
Lots 095112 and 474660.