Z-0528-2008

Baxter Renal Software Suite v.3.0 and v.3.1 - Renal Link HD Module; clinical data management software; Baxter...

Baxter Healthcare Renal Div / initiated 2007-11-14 / Terminated / root cause: software

Baxter Healthcare Renal Div began a recall of Baxter Renal Software Suite v.3.0 and v.3.1 - Renal Link HD Module; clinical data management software; Baxter Healthcare Corporation, McGaw Park, Illinois 60085 on . The reason FDA records is "Pre-Treatment/Post-Treatment Report errors: 1) The Heparin Bolus Value is displayed on Pre and Post Treatment Reports for patients with "No Heparin" orders (RenalSoft v2.0, Renal Software Suite v.3.0 and v.3.1), and 2)...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0528-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Renal Div
Product
Baxter Renal Software Suite v.3.0 and v.3.1 - Renal Link HD Module; clinical data management software; Baxter Healthcare Corporation, McGaw Park, Illinois 60085.
Product code
KPF System, Dialysate Delivery, Semi-Automatic, Peritoneal
Reason for recall
Pre-Treatment/Post-Treatment Report errors: 1) The Heparin Bolus Value is displayed on Pre and Post Treatment Reports for patients with "No Heparin" orders (RenalSoft v2.0, Renal Software Suite v.3.0 and v.3.1), and 2) Withheld Medications are incorrectly displayed as Administered on the Pre-Treatment report (RenalSoft v.2.0 only).
Root cause tag
software
FDA root cause
Software design
Action
Baxter Healthcare Corporation sent an Urgent Device Correction letter, dated 11/14/07, to all of the affected customers, informing them of the Heparin Bolus and the Withheld Medication issues and provided them with interim mitigation instructions that they should implement while waiting for a software correction. A Baxter representative will contact the customer to schedule the software update when the upgraded software is released. Customers were requested to complete and return the enclosed Customer Reply Form, indicating that they have disseminated the information to their staff, and indicating if they still have RenalSoft software and require an upgrade. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898, option 4. A follow-up letter dated 3/6/08 was sent to the consignees on the same date, informing them that the upgraded RenalSoft Software v.2.0.1 was now available for installation. The consignees were requested to complete and fax back the customer reply form indicating that they received the letter, and were informed that they would be contacted by a Baxter representative to schedule an installation of the software upgrade at no charge.
Quantity in commerce
9 units
Distribution
Worldwide Distribution-USA including Puerto Rico, and countries of Australia, Austria, Canada, Chile, Denmark, Finland, Greece, Ireland, Italy, Mexico, The Netherlands, New Zealand, Norway, Panama, Portugal, Spain, Sweden and the United Kingdom
Product codes and lots
Versions 3.0 and 3.1, all Serial Numbers