Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0623-2008

Stryker Orthopaedics; UHR Universal Bipolar Component; 28 x 44 mm, Hip Prosthesis Component, Stryker...

Howmedica Osteonics Corp / initiated 2007-10-26 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics; UHR Universal Bipolar Component; 28 x 44 mm, Hip Prosthesis Component, Stryker Howmedica Osteonics Corp., Mahwah, NJ 07430 on . The reason FDA records is "Incorrect part: Device may have an incorrect locking ring assembled with the UHR bipolar head". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0623-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics; UHR Universal Bipolar Component; 28 x 44 mm, Hip Prosthesis Component, Stryker Howmedica Osteonics Corp., Mahwah, NJ 07430
Product code
KWL Prosthesis, Hip, Hemi-, Femoral, Metal
Reason for recall
Incorrect part: Device may have an incorrect locking ring assembled with the UHR bipolar head.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker sent Urgent Product Hold letters to US branches/agencies and hospitals on October 26, 2007 advising them to quarantine the product from these lots, and Urgent Product Recall letters on November 2, 2007 advising them to return the devices to Stryker. International distributors were sent an advisory on November 6, 2007.
Quantity in commerce
360 total
Distribution
Worldwide Distribution - USA, Canada, China, India, Australia, Korea, Singapore, Taiwan, Sweden, Germany, Netherlands, France, Spain, Italy, Poland, South Africa, and England.
Product codes and lots
Ref No: Description: Lot Code:  UH1-44-28 UHR BIPOLAR 28X44mm 3K4MMD UH1-44-28 UHR BIPOLAR 28X44mm 5P0MLA UH1-44-28 UHR BIPOLAR 28X44mm JEMMNA UH1-44-28 04YMPD UH1-44-28 0YHMMD UH1-44-28 247MMD UH1-44-28 29EMPD UH1-44-28 2HYMRA UH1-44-28 66NMKD UH1-44-28 6W5MPD UH1-44-28 6W6MPD UH1-44-28 8XEMPD UH1-44-28 8XKMPD UH1-44-28 9W0MJA UH1-44-28 D05MKD UH1-44-28 ED8MND UH1-44-28 EP2MPA UH1-44-28 K57MND UH1-44-28 K6LMND UH1-44-28 LA6MPA UH1-44-28 V13MLD UH1-44-28 V17MLD