Z-0551-2008

Abbott, ACCELERATOR Device Manager (ADM); List Number: 08H74-01; Abbott Laboratories, Abbott Park, IL

Abbott Medical / initiated 2007-10-25 / Terminated / root cause: software

Abbott Laboratories, Inc began a recall of Abbott, ACCELERATOR Device Manager (ADM); List Number: 08H74-01; Abbott Laboratories, Abbott Park, IL on . The reason FDA records is "Sample/Patient Mis-identification: Two issues identified related to the ACCELERATOR Decision Manager (ADM), to include: 1) Sample Identification (SID) and/or Patient Identification (PID) numbers that contain more than...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0551-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott, ACCELERATOR Device Manager (ADM); List Number: 08H74-01; Abbott Laboratories, Abbott Park, IL
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Sample/Patient Mis-identification: Two issues identified related to the ACCELERATOR Decision Manager (ADM), to include: 1) Sample Identification (SID) and/or Patient Identification (PID) numbers that contain more than 12 characters are truncated to the final 12 characters by the ADM; 2) for SIDs using alpha characters that are case sensitive, SIDs with lower case characters are transmitted by the
Root cause tag
software
FDA root cause
Software design
Action
Firm notified consignees via Product Correction, Immediate Action Required letter(s) on 10/25/07. Consignees were notified of software issues and informed they would be contacted by an Abbott representative to schedule installation of software upgrade.
Quantity in commerce
19 units.
Distribution
Nationwide:Distributed to hospitals/laboratories in IL, PA, TN, TX, UT, VA and WV.
Product codes and lots
All Serial Numbers.