Z-1156-2008

BIOPLEX 2200 SYPHILIS IgG Kit: 100 tests - Catalog No. 665-1450 - manufactured by Bio-Rad Laboratories in...

Bio-Rad Laboratories Inc. / initiated 2007-11-13 / Terminated / root cause: manufacturing process

Bio-Rad Laboratories began a recall of BIOPLEX 2200 SYPHILIS IgG Kit: 100 tests - Catalog No. 665-1450 - manufactured by Bio-Rad Laboratories in Redmond, WA 98052. The product is a multiplex flow... on . The reason FDA records is "False negative results due to reagent packs exhibiting low signal". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1156-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
BIOPLEX 2200 SYPHILIS IgG Kit: 100 tests - Catalog No. 665-1450 - manufactured by Bio-Rad Laboratories in Redmond, WA 98052. The product is a multiplex flow immunoassay intended for the qualitative detection of Treponema pallidium IgG antibodies in human serum.
Product code
LIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum
Reason for recall
False negative results due to reagent packs exhibiting low signal.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.
Quantity in commerce
796 kits
Distribution
Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.
Product codes and lots
Lot Numbers: 946314 Exp. 10/31/07, 950630 Exp. 1/31/08 and 946899 Exp. 2/28/08.