Z-1157-2008
BIOPLEX 2200 EBV IgG KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1250 - manufactured by...
Bio-Rad Laboratories Inc. / initiated 2007-11-13 / Terminated / root cause: manufacturing process
Bio-Rad Laboratories began a recall of BIOPLEX 2200 EBV IgG KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1250 - manufactured by Bio-Rad Laboratories in Redmond, WA 98052. The product... on . The reason FDA records is "False negative results due to reagent packs exhibiting low signal". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1157-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc.
- Product
- BIOPLEX 2200 EBV IgG KIT, CALIBRATORS, AND CONTROLS: 100 tests - Catalog No. 665-1250 - manufactured by Bio-Rad Laboratories in Redmond, WA 98052. The product is multiflow immunoassay intended for the quantitative detection of IgG antibodies to three (3) seperate EBV antigens in human serum. They are: 1) Epstein-Barr Virus Nuclear Antigen-1 (EBV-NA-1); 2) Viral Capsid Antigen (EBV VCA); and 3) Early Antigen diffuse (EBV EA-D).
- Product code
- LSE Epstein-Barr Virus, Other
- Reason for recall
- False negative results due to reagent packs exhibiting low signal.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.
- Quantity in commerce
- 1,020 kits
- Distribution
- Worldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.
- Product codes and lots
- Lot Numbers: 943710 Exp. 7/30/07 and 943711 Exp. 4/30/08