Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0674-2008

Orthofix TrueLok External Ring Fixation System, Telescopic Linear Distractor, Threaded in the following...

Orthofix SRL / initiated 2007-09-28 / Terminated / root cause: manufacturing process

Orthofix, Inc began a recall of Orthofix TrueLok External Ring Fixation System, Telescopic Linear Distractor, Threaded in the following sizes: 75mm (Catalog Number: 50-1101), 100mm (Catalog... on . The reason FDA records is "Markings reversed: Graduated markings on telescoping bone distractors were reversed, causing the distractors to compress rather than distract the treatment site when adjusted". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0674-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Orthofix TrueLok External Ring Fixation System, Telescopic Linear Distractor, Threaded in the following sizes: 75mm (Catalog Number: 50-1101), 100mm (Catalog Number: 50-1014), 150mm (Catalog Number: 50-1015) and 200mm (Catalog Number: 50-1016), distributed by Orthofix, Inc., McKinney, TX 75069.
Product code
LXT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite
Reason for recall
Markings reversed: Graduated markings on telescoping bone distractors were reversed, causing the distractors to compress rather than distract the treatment site when adjusted.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Orthofix notified sales agents by phone starting on 09/28/07. Sales agents were to return product for replacements.
Quantity in commerce
69 units.
Distribution
Nationwide Distribution - USA states: CA, FL, GA, IL, PA, MI, NY and WI.
Product codes and lots
Lot numbers:  For Catalog #50-1101 (75mm): 075; For Catalog #50-1014 (100mm): 7241, 076 and 50826; For Catalog #50-1015 (150mm): 2664-01, 7343 and 080; For Catalog #50-1016 (200mm): 145, 7342 and 068.