PML -- Vitek NH ID Card (NHI) Set - Quality Control kit for culture media. Each kit includes a 5 pack of...
PML, Inc. dba Pml Microbiologicals Inc. began a recall of PML -- Vitek NH ID Card (NHI) Set - Quality Control kit for culture media. Each kit includes a 5 pack of LyfoCults Haemophilus parainfluenzae - Lot # 185-1 Exp... on . The reason FDA records is "Incorrect micro-organism-- .Quality control In-vitro diagnostic test was manufactured with Cryptococcus neoformans (ATCC 76484) instead of Haemophilus parainfluenzae (ATCC 7901)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0722-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- PML, Inc. dba Pml Microbiologicals Inc.
- Product
- PML -- Vitek NH ID Card (NHI) Set - Quality Control kit for culture media. Each kit includes a 5 pack of LyfoCults Haemophilus parainfluenzae - Lot # 185-1 Exp 13AUG2008. PML Micrcobiologicals, 27120 SW 95th Avenue, Wilsonville, Oregon 97070
- Product code
- JTR Kit, Quality Control For Culture Media
- Reason for recall
- Incorrect micro-organism-- .Quality control In-vitro diagnostic test was manufactured with Cryptococcus neoformans (ATCC 76484) instead of Haemophilus parainfluenzae (ATCC 7901).
- Root cause tag
- manufacturing process
- FDA root cause
- Mixed-up of materials/components
- Action
- On 10/15/2007, firm notified consignees by telephone. A follow up letter dated 10/26/07, was faxed to customers instructing them to discontinue use and discard remaining inventory on site as a biohazardous material. Consignees were requested to fax back response indicating the amount of product they had on hand at time of notification.
- Quantity in commerce
- 1 kit. Kit contains 1 pack (each pack contains 5 vials of Haemophilus parainfluenzae)
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot # 185-1, Catalog # 201130 - Vitek NH ID Card (HHI) Set; Lot # 233851-1, Catalog # 101364 - LyfoCults Microorganism