Z-0526-2008
BD FACS Sample Prep Assistant II, Part Number: 337170; Product is manufactured and distributed by BD...
Becton, Dickinson and Company, BD Biosciences / initiated 2007-10-02 / Terminated / root cause: software
BD Biosciences began a recall of BD FACS Sample Prep Assistant II, Part Number: 337170; Product is manufactured and distributed by BD Biosciences, San Jose, CA 95131 on . The reason FDA records is "Mixing Performance: SPA II devices manufactured with the affected product motor control board may exhibit suboptimal mixing performance". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0526-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- BD FACS Sample Prep Assistant II, Part Number: 337170; Product is manufactured and distributed by BD Biosciences, San Jose, CA 95131
- Product code
- GKI
- Reason for recall
- Mixing Performance: SPA II devices manufactured with the affected product motor control board may exhibit suboptimal mixing performance.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 10/1/07, all consignees were notified via Important Product Information letter and also an on-site visit by the firm's sales representative. The letter informed users of the affected product and provided instructions on the recall. For additional concerns or questions contact BD Customer Support Center at 1-877-232-8995.
- Distribution
- Worldwide Distribution --- USA including states of: NJ, IL, TX, PA, NY, MN, and AZ, and countries of Australia, Belgium, Hong Kong, Trinidad and Tobago. Australia, Belgium, Hong Kong, Trinidad and Tobago.
- Product codes and lots
- Serial Numbers: T0246 to T0290 (except for T0274,T0275,T0277 and T0279)