Z-0846-2008

STERRAD 50 Sterilizer

Advanced Sterilization Products / initiated 2007-11-12 / Terminated / root cause: manufacturing process

Advanced Sterilization Products began a recall of STERRAD 50 Sterilizer, Product Code:10050, Advanced Sterilization Products, Irvine, CA on . The reason FDA records is "User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0846-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Sterilization Products
Product
STERRAD 50 Sterilizer, Product Code:10050, Advanced Sterilization Products, Irvine, CA
Product code
MLR Sterilizer, Chemical
Reason for recall
User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization updated user guidance has been issued.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Urgent Product Correction letters were sent to customers beginning on November 13, 2007 via UPS overnight. Customers were notified this action was being taken to clarify instructions contained within the STERRAD System User Guides to reinforce appropriate use and to address two separate issues which have arisen from user reports: (1) User reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization. (2) User reports about light-colored residues observed on instruments, trays and pouches after sterilization. The Product Correction Letters instructed customers to refer to the revised excerpt of the User Guide (attached to the letter) for information about instrument preparation prior to sterilization, particularly with regard to the washing, rinsing and thorough drying of instruments prior to sterilization in the STERRAD Sterilization System. If customers have any questions they were instructed to call ASP Customer Care Center to answer additional questions at (888) 783-7723.
Quantity in commerce
9,686 units total for recalls Z-0845-0848-2008
Distribution
Worldwide: USA, France, Germany, Italy, The UK, Switzerland, Benelux, Uruguay, Venezuela, Argentina, Chile, Colombia, Ecuador, Belgium, Puerto Rico, Hong Kong, Indonesia, Mexico, The Philippines, Switzerland, Australia, Austria, China, Greece, Ireland, Israel, Japan, Korea, Malaysia, Singapore, Taiwan, Thailand, Turkey, Middle East, Peru, Canada, The Czech Republic, Middle East, Portugal, S Africa, Spain, Sweden, Hungary, Poland, Brazil, Russia, Egypt, India & Slovenia
Product codes and lots
All Serial Numbers